- Place of Origin:
- Shandong, China
- Brand Name:
- BIOBASE
- Model Number:
- BK-UV1000
- Size:
- 560*430*320
- Weight:
- 12 /11.5/ 12/ 11.5
- Focal Length:
- Spectrophotometer
- Wavelength Range:
- 4nm
- Measurement Range:
- 200-1020nm
- Item Name:
- UV/VIS SPECTROPHOTOMETER
- Model:
- BK-UV1000/BK-V1000/BK-UV1200/BK-V1200
- Optical System:
- Single Beam, Grating1200 lines/mm
- Spectral Bandwidth:
- 4nm
- Wavelength Repeatability:
- ≤0.5nm
- Photometric Accuracy:
- ± 0.002A (0-0.5Abs), ± 0.004A (0.5-1.0Abs), ± 0.5% T (0-100% T)
- Photometric Repeatability:
- 0.001Abs (0-0.5Abs), 0.002Abs (0.5-1.0Abs),≤0.2% T (0-100% T)
- Stability:
- ± 0.001A / h @ 500nm
- Noise:
- ± 0.001A
- Photometric Mode:
- T, A, C, E
- Supply Ability:
- 1000 Set/Sets per Month
- Packaging Details
- Wooden case or carton package.
- Port
- Qingdao
- Lead Time:
- Shipped in 10 days after payment
Large LCD Screen UV/VIS Spectrophotometers
BK-UV1000/BK-V1000/BK-UV1200/BK-V1200 Best Price
Application:
Spectrophotometers are widely used for Organic chemistry,Inorganic chemistry,Life sciences,foods,Medicine and health,Agriculture,Geology,Metallurgy and Environment etc.fields.
Features of UV/VIS Spectrophotometers :
Large LCD screen (128*64 Dots)
Imported & Environmental Deuterium lamp, preventing from Ozone inhalation
High-class grating , wholly hermetic light path design, ensures the super low stray light
Real-time monitoring the lifetime of Deuterium lamp and Tungsten lamp with advanced system
Pre-aligned design ensures the user can change lamps conveniently
Wavelength calibration, wavelength setting, change lamp source and dark current calibration automatically
With Si02 coating optical mirror, reducing the pollution from outside fully
Optional PC Software to expand the applications to Quantitative, Multi-Wavelength and Kinetics
Widely used for Organic chemistry, Inorganic chemistry, Life sciences, food, Medicine and health, Agriculture, Geology, Metallurgy and Environment etc. fields
Basic Fuctions:
1. Photometry
Test Abs., Transmittance by fixed wavelength
2. Quantitative
a. C Mode: EstablishC=K*A curve, test sample concentration
b. F Mode: Establish A=K1*C+K0, test sample concentration
c. Display and save curve, test data (V-1200,UV-1200).
Model | BK-UV1000 | BK-V1000 | BK-UV1200 | BK-V1200 | |||
Optical System | Single Beam, Grating1200 lines/mm | ||||||
Wavelength Range | 200-1020nm | 325-1000nm | 200-1020nm | 320-1020nm | |||
Spectral Bandwidth | 4nm | ||||||
Wavelength Accuracy | ±1.8nm | ±1.5nm | |||||
Wavelength Repeatability | ≤0.5nm | ||||||
Photometric Accuracy | ± 0.002A (0-0.5Abs), ± 0.004A (0.5-1.0Abs), ± 0.5% T (0-100% T) | ||||||
Photometric Repeatability | 0.001Abs (0-0.5Abs), 0.002Abs (0.5-1.0Abs),≤0.2% T (0-100% T) | ||||||
Stray Light | ≤0.1 % T @ 360nm; 220nm | ≤0.2% T @ 360nm | |||||
Stability | ± 0.001A / h @ 500nm | ||||||
Noise | ± 0.001A | ||||||
Display | 128*64 Dots LCD | ||||||
Photometric Mode | T, A, C, E | ||||||
Photometric Range | 0-200%T,-0.301~3.0A | ||||||
Detector | Silicon Photodiode | ||||||
Light Source | Deuterium Lamp, Tungsten Lamp | ||||||
Input | Membrane Keypad | ||||||
Standard Accesseries | 10mm glass cuvette *4 units 10mm quartz cuvette *2 units | 10mm glass uvette *4 units | 10mm glass cuvette *4 units 10mm quartz cuvette *2 units | 10mm glass uvette *4 units | |||
Power Supply | 110V/220±10% ,60/50Hz | ||||||
Packing Size(mm) | 560*430*320 | 560*430*320 | 560*430*320 | 560*430*320 | |||
Gross Weight(kg) | 12 | 11.5 | 12 | 11.5 |
Other Model:
BIOBASE GROUP
Established in 1999
Our mission is to enable our customers to make the world healthier,cleanerand safer.
BIOBASE group,serving global distributors and its loyal end users, is becoming one of the most
important scientific services supplier.
Add: NO.51 South Gongye Road, Jinan250101, Shandong, China.