- Type:
- Gas Sterilization Equipments
- Brand Name:
- BIOBASE
- Model Number:
- BKQ-B120II
- Place of Origin:
- Shandong, China
- Instrument classification:
- Class II
- Sterilization Temp.:
- 105-136 degree centigrade
- Chamber Size(mm):
- diameter 386*1055
- Extrnal size:
- 550*640*1370mm
- Capacity:
- 120L
- Chamber material:
- stainless steel
- Working pressure:
- 0.175MPa
- Dynamic pulseexhaust times:
- 0~9 times
- Supply Ability:
- 1000 Unit/Units per Month
- Packaging Details
- Standard exportPackage
- Port
- QingdaoShanghai Tianjin Guangzou
- Picture Example:
There are 5 fixed programs and 4 DIY programs according to
different sterilization objects such as dressings, equipment,
tools, rubber, liquid, culture medium, waste, etc.
Safety Features:
1. Over pressure. Safety valve can automatically open to release pressure.
2. Over temperature. The system will power off automatically.
3. Door sensor. Sensor at door closing position, to check the door status. If door is open, operation will stop.
4. Anti-dry out function. If water level is too low, system will power off and error code displays on screen.
5. Over current. over voltage or electric leakage protection apparatus-air switch.
6. Door cover.The door is covered with safety shell made of thermal insulation material, which can protect operator from burning by heat and looks nice.
Technical features:
1. Complete process of filling water, dynamic pulse exhaust, heating, sterilizing, draining and drying runs automatically.
2. LED screen displays all steps of full process as well as error codes.
3. Sterilization programs for dressings, equipment, rubber, culture medium, waste, etc.
4. Adopting certain temperature (98~102℃) exhaust method and dynamic pulse exhaust method to completely empty the cold air inside chamber for ensuring steam saturability.
5. Automatic selecting quick or low exhaust to avoid sterilization liquid overflowing.
6. Inner stainless steel water tank for water and steam to circulate inside. Steam does not exhaust during sterilization, so as to keep the outside clean and dry.
7. Independent quick service window, easy for repairing and maintenance.
8. Water inlet filter to keep high water quality.
9. Hand wheel type of quick-open door structure, self-expandable sealing ring, easy operating and safe.
10. Optional printer for recording sterilization data.
Parameter:
Model | BKQ-B50(II) | BKQ-B75(II) | BKQ-B100(II) | BKQ-B120(II) | BKQ-B150(II) | BKQ-B200(II) | |
Capacity | 50L | 75L | 100L | 120L | 150L | 200L | |
Chamber size(mm) | φ386×500 | φ386×682 | φ386×875 | φ386×1055 | φ500*760 | φ500*1000 | |
Chamber material | 304 medical stainless steel | ||||||
Max designed pressure | 0.28Mpa | ||||||
Max designed temp | 150℃ | ||||||
Working pressure | 0.175MPa | 0.217MPa | |||||
Working temp. | 105~136℃ | 50~134℃ | |||||
Temp. accuracy | ±0.1℃ | ||||||
Timing range | 0~999min | ||||||
Dynamic pulse exhaust times | 0~9 times | ||||||
Dynamic pulse exhaust temp. | 105~134℃ | ||||||
Dissolution temp. range | 60~100℃(programmable) | ||||||
Temp. holding range | 40~60℃(programmable) | ||||||
Power supply | 110/220V, 60/50Hz | 220V, 50/60Hz | 380V, 50Hz | ||||
Power consumption | 4.4KW | 4.7KW | 5.2KW | 5.2KW | 6KW | ||
SS basket | 1 pc | 2pcs | 2pcs | 3pcs | 3 pcs | 4pcs | |
External size(mm) | 550x640x970 | 550x640x1070 | 550x640x1270 | 550x640x1370 | 715*720*1350 | 715*720*1550 | |
Packing size(mm) | 765*700*1150 | 765*700*1230 | 756*700*1450 | 756*700*1650 | 760*800*1500 | 760*800*1690 | |
Gross weight | 110kg | 114kg | 120kg | 140kg | 180kg | 200kg |
Our service :
Pre-sale service:
24hours online to ensure quick response to customers" inquiry or online message;
Guide customers to choose the best suitable machine model;
Offer detailed specification and competitive prices;
Contract execution:
Machine production will be arranged strictly according to the contract;
Machine photos will be sent to you for inspection during production;
Update latest news after delivery and prepare customs clearance;
After-sale service:
Broken spare parts (except for wear parts) will be sent for free due to quality problems;
In time response to customers technical problems;
New products update for customers" reference.