Automatic calibration system of titration liquid concentration
1.Dual mode of automatic calibration method which instrument eliminates errors and traditional direct input method
2.Automatic calibration method which instrument eliminates errors
Safety protection system
1.Digestion tube vacant protection
2.Safety door vacant protection
3.Automatic water-replenishing function of steam cylinder
4.Steam overpressure protection
5.Leakage protection
Technical parameters
Model
BKN-987
BKN-987A
Working Process
Adding quantitative sample dilution, alkali liquor and absorption liquid; Automatically finishing distillation, titration, print and liquid-discharging.
Measurement Range
0.1-240mgN
Measurement Sample Amount
Solid<6g liquid<16ml
Measurement time
4-8 minutes
Titration Precision
2.0μl/step
0.5μl/step
Titration Solution Concentration
Direct titration samples with 4 titration concentrations of 0.1mol,0.5 mol and 1 mol
Titration System
High-precision plunger type titration system
High-precision peristaltic titration system
Executive Standard
Comply with stipulations of AOAC, EPA, DIN, ISO AND GB
Repeatability Precision
±0.5%
Recovery Ratio
≥99.5%
Data Storage
Effective storage space 2G
Print Function
110mm line printer which can realize user selective print function
Power
1800w
Power Supply
AC220v, 50/60Hz
Requirement of Cooling Water
Under 25℃
Cooling Water Consumption
1.5L/min (Distillation process)
Packing Size(W∗D∗H)
510∗460∗710mm
Gross Weight
36.4kg